User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Kyowa Frequency Response Sub DATE: 27-Aug-97 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 19 NUMBER OF LINES: 117 CONFIGURATION: Fluke 5700A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 MEME <1 1.002 DISP Connect the Fluke 5700 output to the UUT CH [mem] 1.002 DISP RECORD BNC. Connect the UUT CH [mem] REPRODUCE BNC 1.002 DISP to the 34401A HI/LO INPUT. 1.003 DISP Set the UUT TAPE SPEED to 38 cm/s. 1.004 DISP On the UUT depress the REC and FWD buttons 1.004 DISP simultaneously. 1.005 HEAD { 38 cm/s } 1.006 HEAD Frequency Response Tests 1.007 JMP 2.001 1.008 EVAL 2.001 MESS Note the UUT % meter reading. 2.002 5700 0D 100H N 2W 2.003 MEME 2.004 MEMI Enter the UUT % METER reading. 2.005 MEME 2.006 MESS Adjust the stimulus for the [mem1]% reading on the 2.006 MESS UUT METER. 2.007 5700 0.000D -1U +0.5U 10H C 2W 3.001 5700 0.000D -1U +0.5U 10kH C 2W 4.001 MATH m[1] = -49 4.002 CALL Signal to Noise Sub 4.003 DISP Set the UUT TAPE SPEED to 19 cm/s. 4.004 HEAD { 19 cm/s } 4.005 HEAD Frequency Response Tests 4.006 JMP 5.001 4.007 EVAL 5.001 MESS Note the % reading on the UUT METER. 5.002 5700 0D 100H N 2W 5.003 MEME 5.004 MEMI Enter the UUT % METER reading. 5.005 MEME 5.006 MESS Adjust the stimulus for the [mem1]% reading on the 5.006 MESS UUT METER. 5.007 5700 0.000D -1U +0.5U 10H C 2W 6.001 5700 0.000D -1U +0.5U 5kH C 2W 7.001 MATH m[1] = -48 7.002 CALL Signal to Noise Sub 7.003 DISP Set the UUT TAPE SPEED to 9.5 cm/s. 7.004 HEAD { 9.5 cm/s } 7.005 HEAD Frequency Response Tests 7.006 JMP 8.001 7.007 EVAL 8.001 MESS Note the % reading on the UUT METER. 8.002 5700 0D 100H N 2W 8.003 MEME 8.004 MEMI Enter the UUT % METER reading. 8.005 MEME 8.006 MESS Adjust the stimulus for the [mem1]% reading on the 8.006 MESS UUT METER. 8.007 5700 0.000D -1U +0.5U 10H C 2W 9.001 5700 0.000D -1U +0.5U 2.5kH C 2W 10.001 MATH m[1] = -46 10.002 CALL Signal to Noise Sub 10.003 DISP Set the UUT TAPE SPEED to 4.8 cm/s. 10.004 HEAD { 4.8 cm/s } 10.005 HEAD Frequency Response Tests 10.006 JMP 11.001 10.007 EVAL 11.001 MESS Note the % reading on the UUT METER. 11.002 5700 0D 100H N 2W 11.003 MEME 11.004 MEMI Enter the UUT % METER reading. 11.005 MEME 11.006 MESS Adjust the stimulus for the [mem1]% reading on the 11.006 MESS UUT METER. 11.007 5700 0.000D -1U +0.5U 10H C 2W 12.001 5700 0.000D -1U +0.5U 1.25kH C 2W 13.001 MATH m[1] = -44 13.002 CALL Signal to Noise Sub 13.003 DISP Set the UUT TAPE SPEED to 2.4 cm/s. 13.004 HEAD { 2.4 cm/s } 13.005 HEAD Frequency Response Tests 13.006 JMP 14.001 13.007 EVAL 14.001 MESS Note the % reading on the UUT METER. 14.002 5700 0D 100H N 2W 14.003 MEME 14.004 MEMI Enter the UUT % METER reading. 14.005 MEME 14.006 MESS Adjust the stimulus for the [mem1]% reading on the 14.006 MESS UUT METER. 14.007 5700 0.000D -1U +0.5U 10H C 2W 15.001 5700 0.000D -1U +0.5U 625H C 2W 16.001 MATH m[1] = -42 16.002 CALL Signal to Noise Sub 16.003 DISP Set the UUT TAPE SPEED to 1.2 cm/s. 16.004 HEAD { 1.2 cm/s } 16.005 HEAD Frequency Response Tests 16.006 JMP 17.001 16.007 EVAL 17.001 MESS Note the % reading on the UUT METER. 17.002 5700 0D 100H N 2W 17.003 MEME 17.004 MEMI Enter the UUT % METER reading. 17.005 MEME 17.006 MESS Adjust the stimulus for the [mem1]% reading on the 17.006 MESS UUT METER. 17.007 5700 0.000D -1U +0.5U 10H C 2W 18.001 5700 0.000D -1U +0.5U 313H C 2W 19.001 MATH m[1] = -41 19.002 CALL Signal to Noise Sub 19.003 DISP Press the STOP button on the UUT. 19.004 MESS 19.005 END